Cookies ussage consent

Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.

A new time-gated synchronization controller for the detection of singlet oxygen luminescence

TONGQING HUANG1, DEFU CHEN1, WEIDA SU1, HUIYUN LIN1, BUHONG LI1,* , SHUSEN XIE1

Affiliation

  1. Key Laboratory of OptoElectronic Science and Technology for Medicine of Ministry of Education, Fujian Provincial Key Laboratory for Photonics Technology, Fujian Normal University, Fuzhou 350007, P. R.

Abstract

A new time-gated synchronization controller is developed for the detection of near-infrared singlet oxygen (1O2) luminescence around 1270 nm, in which the position and width of the sampling gate can be adjusted according to the time-resolved characteristics of 1O2 luminescence in different environments, respectively. Using this controller, the short-lived and long-lived background luminescence can be efficiently suppressed, which dramatically improves the signal-to-noise ratio for the measurement of time-integrated 1O2 luminescence. This new controller has the potential application for the detection of 1O2 luminescence.

Keywords

Singlet oxygen, Luminescence, Detection, Time-gated synchronization controller, Signal-to-noise ratio.

Citation

TONGQING HUANG, DEFU CHEN, WEIDA SU, HUIYUN LIN, BUHONG LI, SHUSEN XIE, A new time-gated synchronization controller for the detection of singlet oxygen luminescence, Optoelectronics and Advanced Materials - Rapid Communications, 4, 8, August 2010, pp.1083-1087 (2010).

Submitted at: June 21, 2010

Accepted at: Aug. 12, 2010