Our site saves small pieces of text information (cookies) on your device in order to deliver better content and for statistical purposes. You can disable the usage of cookies by changing the settings of your browser. By browsing our site without changing the browser settings you grant us permission to store that information on your device.

Continuous trace gas concentration measurement based on quartz-enhanced photoacoustic spectroscopy technique

Y. TONG1, Y. F. MA1,*

Affiliation

  1. National Key Laboratory of Science and Technology on Tunable Laser, Harbin Institute of Technology, China

Abstract

A continuous trace gas concentration measurement using quartz-enhanced photoacoustic spectroscopy technique based on harmonic detection and LabVIEW program was demonstrated in this paper. The wavelength of the laser was locked at the absorption line peak with the help of the third harmonic signal. Using a 2.33 μm continuous wave, distributed feedback fiber-coupled diode laser, a long-term monitoring of carbon monoxide concentration for 85 s was performed to demonstrate reliable operation of the quartz-enhanced photoacoustic spectroscopy sensor..

Keywords

Continuous gas concentration measurement, QEPAS, Harmonic detection, Absorption line locking.

Citation

Y. TONG, Y. F. MA, Continuous trace gas concentration measurement based on quartz-enhanced photoacoustic spectroscopy technique, Optoelectronics and Advanced Materials - Rapid Communications, 13, 1-2, January-February 2019, pp.51-55 (2019).

Submitted at: May 16, 2018

Accepted at: Feb. 12, 2019